Latest addition to the Cr - series product family is the high sensitivity Cr100m FTIR - gas analyser with a 100 meter sample cell. The detection limits of the Gasmet™ Cr100m analyser are as low as 10 ppb (parts per billion). Applications of Cr100m include quality control in high purity gas manufacturing and industrial hygiene measurements.

Multicomponent FTIR Gas Analyzer
GASMET IN-LAB SERIES includes quantitative multicomponent gas analyzers for laboratory research applications. The GASMET Cr-100 incorporates a Fourier Transform Infrared, FT-IR spectrometer, ambient temperature sample cell, and signal processing electronics. Liquid nitrogen cooled MCT detector has the highest performance available. The GASMET Cr-100 is designed for highest sensitivity and speed. The long path length of the sample cell provides low detection limits. The mirrors are gold coated and the sample cell is not heated.
General parameters
Measuring principle: Fourier Transform Infrared, FTIR
Performance: Simultaneous analysis of up to 50 gas compounds
Response time, T90: Depends on the gas flow and measurement time
Operating temperature: 15 - 25°C non condensing
Storage temperature: -20 - 60°C, non condensing
Power supply: 100-115 or 230 V / 50 -60 Hz
Power consumption: Max 100 W
Spectrometer
Resolution: 8 cm-1 or 4 cm-1
Scan frequency: 10 scans / s
Detector: Liquid N2 cooled MCT
Source: SiC, 1550 K
Optics material: ZnSe
Wavenumber range: 600 - 4200 cm-1
Sample Cell
Structure: Multi-pass, fixed path length 100 m
Material: Aluminium
Mirrors: Fixed, reflective gold coating
Volume: 30 l
Connectors: Swagelok 16 mm
Gaskets: Viton® O-rings
Temperature: Ambient
Window material: ZnSe
Measuring parameters
Zero point calibration: 24 hours, calibration with nitrogen (4.0 or higher N2 recommended)
Zero point drift: < 2 % of measuring range per zero point calibration interval
Sensitivity drift: None
Linearity deviation: < 2 % of measuring range
Temperature drifts: < 2 % of measuring range per 10 K temperature change
Pressure influence: 1 % change of measuring value for 1 % sample pressure change. Ambient pressure changes measured and compensated
Electrical Connectors
Digital Interface: 9-pole D-Connector for RS-232
Cr-100 is connected to an external computer via RS-232C cable. The external computer controls the GASMET.
Power connection: Standard plug CEE-22
Gas Inlet and Outlet Conditions
Gas temperature: Non-condensing, the sample gas temperature should be the same as the sample cell temperature
Flow rate: 120 - 600 l per hour
Gas filtration: Filtration of particulates (2μ) required
Sample gas pressure: Ambient
Sample pump: External, not included
Electronics
A/D Converter: Dynamic range 95 dB
Signal Processor: 32-bit floating point DSP 120 MFLOPS
Computer: External, not included
Analysis Software (for external PC)
Operating system: Windows XP
Analysis software: CALCMET for Windows
Enclosure
Material: Aluminium
Dimensions (mm): 1225 * 450 * 311
Weight: 45 kg
CE - Label: according to EMI guideline 89/336/EC