Artikel/Pagina Bijlagen (0) Artikel (0) Informatie aanvraag
Large Scale Calorimeter

Large-volume calorimeter for characterizing nuclear materials and waste

 

 

Advantages include:
Independent of the matrix.
Low cost compared with other methods, no maintenance costs
Non-destructive assay (NDA).
Possible coupling with gammametry.
General Principle
The calorimeter is of the twin-cell symmetrical heat flow type. It consists of a thermostat-controlled block, opening at the top and surrounding two wells (cavities), each designed to receive a container. One cavity the sample or measuring chamber is loaded with an active product, while the other the
reference chamber is left empty or loaded with an inactive substance.
The six surfaces of each cavity are covered with flowmetric plates (Peltier elements) that can measure an electrical signal (e.m.f.) proportional to the heat flow between the interior of the chamber and the thermostated block.
As the Peltier elements cover all six surfaces, virtually all the heat flow produced in the chamber and transmitted to the exterior can be integrated.

Our Large Scale Calorimeters are available in different versions:
- 15 liters: Series 270
- 60 liters: Series 390
- 90 liters: Series 680
- and up to 200 liters on special request

Large-volume calorimeter for characterizing nuclear materials and waste

 

 

 

 

 

 

DIMENSIONS

 

Measure and reference cells

Series

270

390

680

Special

Internal volume

15 L

60 L

90 L

up to 200 L

 

General Dimensions

height/Width/Depth (in mm)

850/1040/780

960/1500/1000

1260/1500/1000

NA

Mass (kg)

500

1000

1200

NA

 

PERFORMANCES

Limit of detection

260W

500W

1000W

NA

Measuring range

0.015 - 3000mW

0.5-13000 mW

1-26000 mW

NA

Sensitivity

80V/mW

160V/mW

155V/mW

NA

Signal stabilization time

6 hours

30 hours

NA

Working temperature

isothermal

NA

 

 

 


Copyright: Ankersmid 2009 Contact: info@ankersmid.com