Artikel/Pagina Bijlagen (1) Artikel (0) Informatie aanvraag
Rheometer voor Reologische analyse

Rheometer RC20

for quality control, research and development

The Rheometer RC20 is a rotational viscometer (Searle principle) for quality control as well as for product development and engineering research in the laboratory.

The highly dynamic measuring drive system enables rotational tests with controlled shear rate or controlled shear stress (CSR and CSS).

Special features

  • microprocessor-controlled measuring device for mains power operation
  • torque overload protection
  • PC controlled measuring by using Rheometer Software RHEO2000 for graphical data-presentation and regression
  • user-friendly, dust proof keyboard for manual use
  • integrated alphanumeric display (4 lines) for optional displaying of current parameters of viscosity, shear rate, shear stress, rotational speed, torque, temperature
  • data buffering in RAM, storage of measuring data with and measuring-count for up to 10 measurements with up to 4,800 data-points
  • assortment of measuring systems and attachments

Technical data

torque range mains power

50 mNm

torque resolution

0.01 mNm

angle resolution*

15.7 µrad

speed range

0.1 to 1000 rpm

viscosity range**

10-3 to 3 * 103 Pas

shear rate range**

0.1 to 4 * 103 s-1

shear stress range**

0.7 to 3.4 * 104 Pa

accuracy

± 1% of maximum value

temperature range

-20 °C to +180 °C

temperature sensor

Pt 100 1/3 DIN

power supply

100 to 240 V alternating current, 20 W

* high resolution option
** depending on measuring system used

 

Optional Test Mode

  • CSR - rotational test with shear rate control (manually and via PC software)
  • shear rate steps and ramp-functions with defined speeds or shear rates
  • CSS - rotational test with controlled shear stress (manually and via PC software)
  • shear stress levels and ramp-functions with defined torque or shear stress
  • creep test

Measurable and Evaluable Data

  • viscosity
  • shear rate
  • shear stress
  • speed
  • torque
  • temperature
  • time

Attachments

  • thermostatic device for cylinder measuring systemes
  • temperature sensor Pt100
  • standard cylinder measuring systems (DIN 53019), also as one way measuring system
  • double gap measuring systeme (DIN 54453)
  • optional measuring device for cone/plate and plate/plate measuring systems
  • vertically adjustable stand

Additional equipment

  • PC (minimum configuration - Pentium II, 400 MHz, 64 MB RAM, Windows 95)
  • liquid circulator
  • viscosity standard or calibration fluid

Memory: 1548Copyright: Ankersmid 2009 Contact: info@ankersmid.com