Measuring
Carbon dioxide CO2
Carbon monoxide CO
Hydrogen chloride HCl
Hydrogen fluoride HF
Methane CH4
Nitric oxide NO
Nitrogen dioxide NO2
Nitrogen oxides NOx
Nitrous oxide formerly nitrogen oxide N2O
Oxygen O2
Sulfur dioxide SO2
Merk: Envea
Techniek:
IR
Type meting:
In-situ
Complete QAL 1 systeem MIR IS
country
Request information
Request information for: Complete QAL 1 systeem MIR IS

In-situ Close-coupled Multi-Gas Analyzer MIR IS
A complete “all in one compact” system, for multi-gas measurements, based on the field-proven MIR 9000 analyzer and on-board SEC sampling system
Fast & simultaneous measurement of up to 10 gases among: HCl, NO, NO2 (NOx), SO2, CO, CO2, HC, CH4 (TOC), HF, N2O, O2, at the sampling location
• Integrated sample drying & system conditioning; no sample line necessary
Features & Benefits
- Robust analyzer with a stainless steel enclosure
- Designed for measuring wet & corrosive samples
- Ease of installation (single stack entry, on-stack or close-coupled) for reduced costs
- Flow, Temperature & Pressure parameters (optional)
- Remote control and display functions
- Graphic LCD screen with real time synoptic
- Intrinsic security with on-board residual H2O measurement
- On-board paramagnetic cell for long term O2 measurement
- Fast response time, with automatic cross interference correction
- Robust and reliable short extractive analyzer with built-in permeation sample drying system for the measurement of wet and corrosive sample
- Complete CEMS in one cabinet – combines infrared analyzer, sample conditioning
- system and probe into one device
- Highly accurate; excellent calibration stability with automatic optical check
- Heated probe in different materials and lengths to suit the application
- External signal input for up to 5 additional measurements (flow rate, pressure, gas temperature…) or any other analog input
Main applications
- Industrial Boilers and Furnaces, Chemical and Petrochemical Plants
- Incineration, Co-generation, Gas Turbines, Power Plants
- Process Control, Pre/post Scrubber Measurements
- Coal / Oil / Gas Fired Combustion Monitoring
- CHP Combustion Monitoring
| Measurable parameters | |||
|---|---|---|---|
| Pollutant | Range | ||
| HCI | 0-15 / 5000 | ||
| HF | 0-20 / 300 | ||
| NO | 0-100 / 5000 | ||
| NO2 | 0-100 / 1000 | ||
| N2O | 0-20 / 1000 | ||
| NOx | 0-200 / 5000 | ||
| SO2 | 0-75 / 5000 | ||
| CO | 0-75 / 5000 | ||
| CH4 | 0-10 / 1000 | ||
| THC | 0-50 / 5000 | ||
| CO2(%) | 0-10 / 100 | ||
| O2(%) | 0-10 / 25 | ||
Lowest / Highest available ranges expressed in mg/m3 (may vary with your site conditions to be indicated on the Site Survey Form)
| Technical Specifications | |
|---|---|
| QAL1 to EN 15267-3 certified | YES |
| US EPA compliance | YES |
| Sampling technology | In-Situ |
| Sample transfert | N/A |
| back flush function | YES |
| span injection | YES at sample point |
| Built-In O2 option | YES |
| Serial link communication | RS232 / RS422 |
| Other Communication | Ethernet |
| Analogue OUTPUTS | YES (x 8) |
| Analogue INPUTS | YES (x 4) |
| Dimensions (mm) | SS Tight box: 200 x 600 x 600 (D x W x H) |
Approvals & Certifications
- COMPLIANCE WITH WID AND LCPD APPLICATIONS
- QAL 1 certified by the MCERTs to EN 15267-3
- MCERTs QAL 1 certified gas monitor
- Sira MC 040030/09
- QAL3 compliance as defined by EN 14181
- U.S. EPA approved as compliant with 40 CFR 60 ET 75
envea_catalogue_cems_emission-monitoring_en